All too frequently, sophisticated and well-equipped
microscopes fail to yield excellent images because of incorrect use of the
light source. Excellent illumination of the specimen should be bright,
glare-free, and evenly dispersed in the field of view. Since most modern
microscopes achieve such excellence by us of Koehler illumination
(so-called after its discoverer, August Koehler), this description will
deal with the method to achieve Koehler illumination.
There are several physical-mechanical requirements. The
sub-stage condenser must be capable of being focused up and down,
preferably by a knob operating on a rack and pinion. The sub-stage
condenser must be fitted with an aperture iris diaphragm that can be
opened and closed by a lever or knob. The lamp must be fitted with a
condensing lens, a collector, and a field iris diaphragm that can be
opened and closed. It is also desirable that the lamp filaments or bulb be
centerable or pre-centered. To repeat, there are two important adjustable
iris diaphragms; the aperture diaphragm at the sub-stage condenser and the
field diaphragm nearer to the lamp. The aperture iris diaphragm controls
the angular aperture of the cone of light from the condenser. The field
iris diaphragm controls the area of the circle of light illumination the
specimen.
Step 1. Open the aperture
iris diaphragm wide and also open the field iris diaphragm wide.
Turn on the lamp. Using a low power objective (10x or so) and a 10x
eyepiece, slowly focus the specimen that has been placed on the microscope
stage over the stage opening.

Step 2. Close the field
iris diaphragm most of the way. Using the rack and pinion knob of the
condenser, raise the condenser until the edges of the partly closed field
iris diaphragm appear superimposed on the already-focused specimen. The
edges of the field iris diaphragm should appear sharply focused.
The substage condenser is usually close to its highest position.
Step 3. If the field
diaphragm does not appear centered in the field of view, use the substage
condenser centering screws to center the field diaphragm. Then
slowly open the field iris diaphragm until it just disappears from
view. This step must be repeated each time a different objective is turned
into place on the nosepiece.
Step 4. Now lift the eyepiece
out of the body tube and look down the tube at the back lens of the
fully-lighted objective. (This is best accomplished by the use of a
pinhole eyepiece – an eyepiece with a tiny hole but no lenses or a
focusing telescope such as is provided for use in phase – contrast
microscopy.) while looking down the microscope tube, slowly open and close
the substage condenser aperture iris diaphragm. It will be seen
that closing the aperture iris diaphragm "cuts into" the
periphery of the back lens of the objective. For excellent illumination
and contrast, approximately ¼-1/3 of the back lens should be occluded,
thus leaving 2/3-3/4 of the back lens illuminated. Then replace the
eyepiece in the tube. This step too must be repeated each time a different
objective is turned into place on the nosepiece.
The aperture iris diaphragm may have a calibrated scale
(sometimes the calibration refers directly to numerical aperture) which
can be used to make the iris adjustment more accurate as well as readily
repeatable.
It will now be found that the specimen iris
well-illuminated with even, glare-free light, giving good image contrast.
The intensity of the lamp can be adjusted by proper use of the transformer
or by neutral density filters, not by raising or lowering the
condenser, not by closing the aperture iris diaphragm.
Increasing the voltage of the transformer controlling the lamp increases
the color temperature of the light. Such color temperature must be
properly adjusted in color photomicrography because different color films
are balanced for different color temperatures of the light source.