            Products Contact IPS Support and Services


Animation of raw and deconvolved Z-stack images of a pollen grain.
On the right, pollen has been deconvolved using SharpStack's nearest neighbor algorithm.
(Animation features are part of Image-Pro Plus.)



One frame with an AOI selected in a Z-stack on the left with two AOIs and their deconvolved counterparts. Click on image for more detail

SharpStackTM
The Deconvolution Module for Image-Pro Plus

Obtaining clear images from a Z-stack has been a challenge for microscopists. The SharpStack add-on module for Image-Pro Plus meets the challenge by extracting clear, sharp images from a stack of hazy planes.

Unlike other deconvolution solutions, SharpStack integrates seamlessly with image capture, 2D processing, analysis and reporting within the world's most popular image analysis software, Image-Pro Plus.

With SharpStack, nearest neighbor, no neighbor, and inverse algorithm functions are employed to easily sharpen one or all planes from a Z-stack.

 


Pig cerebellum image deconvolved using the inverse filter algorithm.

Elucidating 3D Structures

Cells and tissues are three-dimensional structures, so the observed image at the focal plane contains information from the plane on which the microscope is focused as well as "out-of-focus" contributions from other parts of the specimen above and below the plane of focus. Due to these "out-of-focus" contributions, interpretation of the 3D structure of the biological specimen is hampered in optical imaging. The image at a given focal plane is a poor representation of a true section through the thick specimen. To remedy these problems and produce more reliable 3D data, two methods are widely employed: confocal microscopy and digital deconvolution.

Confocal Microscopy

A confocal microscope uses the pinhole aperture to restrict the "out-of-focus" flare reaching a single detector, the photomultiplier tube (PMT). Because of the pinhole aperture, all the fluorescence signal collected by the objective lens is not used. In order to increase the signal, increased excitation laser intensity is used, but this will often introduce photobleaching and photodamage. Another disadvantage of confocal microscopy is its relatively high cost.

Why Use Digital Deconvolution?

Unlike the confocal configuration described above, digital deconvolution microscopy uses the entire fluorescence signal collected by the objective lens without using pinhole to deliver the emitted light to 2D high-sensitivity CCD cameras. The "out-of-focus" flare introduced into the imaging at different optical sections with its subsequent image degradation is reversed by computer deconvolution through the use of a pointspread function (PSF) of the imaging system! By modeling the microscope optics as a linear and shift-invariant system, the PSF can be used to describe the transformation of any image by the microscope. A typical fluorescence microscope image can be:

[measured image] = [PSF] * [desired image]
where '*' symbol represents the mathematical operation of convolution. The deconvolution (*-1), the mathematical inverse of convolution can be represented as:

[desired image] = [measured image] (*-1) [PSF] The goal of deconvolution is to solve the equation for the desired image.

Digital Deconvolution Methods

The Inverse Filter is a one-step non-iterative approach based upon inverse filtering theory. The deterministic blurring as a convolution of the image with the point spread function can be modeled. In the frequency domain a convolution transforms into a multiplication of the Fourier transform of the sample with the optical transfer function. The optical transfer function (OTF) is the Fourier transform of the point spread function. The inverse filter then accomplishes image restoration by dividing the Fourier transform of the image by the OTF.

The Nearest Neighbor algorithm works by deconvolving one image slice at a time. It uses information from image slices that reside above and below the image slice that is being processed. The precise increment slice position from the measured slice is user selectable. If the slices are chosen judiciously, this approximation will produce results very close to the inverse filter method but takes much less time.

The No Neighbor method uses the information from each single slice to construct a 2D PSF. This is the fastest but may not be as representative of the sample as the other methods.

Inverse filter, nearest neighbor, and no neighbor algorithm functions are all included in SharpStack.

SharpStack System Requirements

bulletImage-Pro Plus version 4.5 or higher.
bulletPentium III CPU, running at 450 Mhz or higher
bulletMicrosoft Windows™ 98/ME/NT/2K
bullet256 MB of RAM (512 MB recommended)
bullet3 GB disk drive with available disk space for installation with additional space to accommodate four times the image size with four bytes/pixel
bulletColor monitor displaying 16-bit high color (24 or 32- bit color preferred)

 

For further information, or a product demonstration, please feel free to contact IPS.

 